Telescopic probe for 6150 AD 5/6 for measuring X-ray and gamma radiation, and detecting Beta radiation

  • Automatically switches between 2 GM tubes to cover wide dose range.
  • Approved for use on Magnox sites.
  • Safety through distance: Extends continuously to approx. 4.1m
  • Robust waterproof aluminium housing design.

The telescopic probe of the teletector allows an operator to measure in places that are difficult to reach, as well as increasing the distance to the radiation source and reducing their exposure accordingly.

   
Low Range Detector Beta Gamma end window tube ZP1400 or equivalent, effective length 40 mmm sensitivity approx. 5800 pulses per µSv
High Range Detector Gamma tube ZP1300 or equivalent, effective length 8 mm, sensitivity approx. 100 pulses per µSv
Energy Range and Angular Range Max. ±40% within:Low range tube: 65 keV - 1.3 MeVHigh range tube: 65 keV - 3 MeV and ±45°(preferential direction is perpendicular to the axis)
Temperature -30°C to +50°C, deviation max. ±10% referred to indication at +20°C
Humidity Nominal range 0 - 95% within specificied temperature range
Power Supply 4.75 V out of 6150 AD
Housing Aluminium, die-cast, scratch resistant paint, waterproof, protection class IP 67 according to DIN 40050 if telescope pushed together and protective cap applied
Dimensions Length 910 - 4120 mm depending on telescope positionWidth 130 mm, Height 90 mm
Weight Approx. 3 kg excluding the 6150 AD

Documents

The LabLogic Group will always treat your details with the utmost care and will never sell them to third party companies.

Please choose one of the following:



    Previous
    Next